It has been reported that on (b)(6) 2019 at around 2pm a noticeable increase of the breathing rate was noticed.Subsequently, the screen was white and therefore no longer operable.The user then pressed the green on/off button and the device then returned to standby mode.From there, it was easy to switch back into a ventilation mode and to continue the ventilation without any problems.No injury reported.
|
A similar error pattern has already been reported to us from the same hospital (kantonsspital graubünden).In this context, the affected device was investigated by the manufacturer and was basis for the investigation, in addition to the analysis of the devices log books and an assessment of the situation on-site.Based on the log analysis the occurrence of reboots could be comprehended.Before the reboots, there were indications of increasing processor utilization which may have resulted in a delayed display of real-time curves and therapy settings on the display, possibly causing a temporal compression of co2, pressure and flow curves, temporarily giving the impression of an increased breathing rate.The reported increased breathing rate may also have been caused by spontaneous breathing activities independently of the reported error pattern, since in autoflow the synchronization was switched on.During the on-site investigation, the reported failure reoccurred.It could be determined that a loose contact in the connection cable (not dräger) of the connected philips monitors was the cause of the failure.Further investigations were performed in the laboratory.It could be shown that the error mechanism could be provoked by a short-term interruption of the receiving line of the used cable.Therefore, it can be summarized that a faulty cable (connecting the philips monitors) was the root cause of the reported symptoms.As a remedy, the ribbon cables used by the customer were replaced by more robust round cables.
|