• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ARCOS 24X250MM SPL TPR DIST; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ARCOS 24X250MM SPL TPR DIST; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Multiple mdr's were filed in association with this reporting: 0001825034-2019-05188, 0001825034-2019-05189, 0001825034-2019-05190, 0001825034-2019-05191, 0001825034-2019-05192.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported during the distribution process packaging damage with sterility barrier potentially compromised was identified.No patient or surgical involvement.No further information available at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported packaging damage event was confirmed via visual examination.It was identified the packaging was damaged in transit.Device history record (dhr) review was performed with no related manufacturing deviations or anomalies identified.Root cause is attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information available at the time of this reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCOS 24X250MM SPL TPR DIST
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9334517
Report Number0001825034-2019-05188
Device Sequence Number1927195
Product Code KWY
UDI-Device Identifier00880304474550
UDI-Public(01)00880304474550
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K090757
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberN/A
Device Catalogue Number11-301024
Device Lot Number917190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/30/2019
Supplement Date Received by Manufacturer01/22/2020
Initial Report FDA Received Date11/18/2019
Supplement Report FDA Received Date01/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-