• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JETSTREAM SC; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JETSTREAM SC; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 45006
Device Problems Entrapment of Device (1212); Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/30/2019
Event Type  malfunction  
Event Description
It was reported that the device became stuck on the wire.A 1.85mm jetstream atherectomy catheter was selected use for an atherectomy procedure in the superficial femoral artery (sfa).During the procedure, the jetstream catheter was advanced over a jetwire guidewire to 100% stenosed, minimally tortuous and less than 50% calcified target lesion.The catheter could not cross over the tightly calcified lesion.Then, the jetstream device became stuck in the wire.The catheter and the wire were removed together from the patient then the jetstream catheter came out freely.The procedure was completed with angioplasty balloon and stent.There were no patient complications reported and the patient's status was fine post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM SC
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9335194
MDR Text Key173450626
Report Number2134265-2019-13924
Device Sequence Number1
Product Code MCW
UDI-Device Identifier08714729889861
UDI-Public08714729889861
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/19/2021
Device Model Number45006
Device Catalogue Number45006
Device Lot Number0023368542
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/30/2019
Initial Date FDA Received11/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/19/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age54 YR
-
-