ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1050-020
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2019001.The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was performed to treat a non-calcified, non-tortuous left forearm fistula.The 5.0x20mm armada 35 balloon failed to deflate.After 3-5 minutes of attempting to deflate, the balloon partially deflated and was removed.A new unspecified balloon was used to successfully continue the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Without having the device to examine, the investigation was unable to determine a conclusive cause for the reported deflation difficulty.D10, h3 - it was initially reported that the device would be returned for analysis.Subsequent information revealed that the device was discarded and is not available for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9335517
Report Number2024168-2019-13674
Device Sequence Number19295222
Product Code LIT
UDI-Device Identifier08717648154591
UDI-Public08717648154591
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2020
Device Model NumberB1050-020
Device Catalogue NumberB1050-020
Device Lot Number70505G1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/04/2019
Supplement Date Received by Manufacturer01/07/2020
Initial Report FDA Received Date11/18/2019
Supplement Report FDA Received Date01/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-