• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT POR TAPER SZ2 STD OFF; SUMMIT HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT POR TAPER SZ2 STD OFF; SUMMIT HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1570-01-080
Medical Device Problem Codes Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Bone Fracture(s) (1870); Hypersensitivity/Allergic reaction (1907); Test Result (2695); No Code Available (3191)
Date of Event 10/22/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the surgeon replaced metal liner with poly liner and ts ceramic head due to elevated ion levels.Bilateral metal-on-metal hip patient was asymptomatic prior to falling and fracturing left hip in inter-trochanteric area.The surgeon had planned to place a cable around the inter-trochanteric region of the left femur but ultimately decided that it wasn't necessary.Implants remaining in patient: summit porous size 2 std; 50mm pinnacle sector cup; 35mm x 6.5mm acetabular screw.Doi: (b)(6) 2005; dor: (b)(6) 2019; left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate no code available (3191) is used to capture fall, blood heavy metal increased and surgical intervention.
 
Event or Problem Description
After a review of the medical records, the patient was revised to address the left proximal fracture on the femur, osteolysis, metal ion toxicity, pseudotumor, and minimal taper wear on the femoral stem.Revision note indicated that there was a large amount of resolving hemarthrosis fluid that was found after t-capsulotomy was performed.The head has minimal taper wear and small localized osteolytic on the cup.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.H10 additional narrative: added: b1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT POR TAPER SZ2 STD OFF
Common Device Name
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9336858
Report Number1818910-2019-116711
Device Sequence Number1873458
Product Code JDI
UDI-Device Identifier10603295059332
UDI-Public10603295059332
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K170339
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1570-01-080
Device Catalogue Number157001080
Device Lot NumberY2FEL1000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/22/2019
Supplement Date Received by Manufacturer12/05/2019
06/16/2020
07/07/2020
Initial Report FDA Received Date11/18/2019
Supplement Report FDA Received Date12/11/2019
07/01/2020
07/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP; APEX HOLE ELIM POSITIVE STOP; ARTICULEZE M 28MM 8.5; PINN CAN BONE SCREW 6.5MMX35MM; PINN CAN BONE SCREW 6.5MMX35MM; PINNACLE BONE SCREW 15 MM; PINNACLE BONE SCREW 15 MM; PINNACLE MTL INS NEUT28IDX50OD; PINNACLE SECTOR II CUP 50MM; PINNACLE SECTOR II CUP 50MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient Weight64
-
-