Model Number MMT-723LNAS
Device Problems Premature Discharge of Battery (1057); Mechanical Problem (1384); Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/02/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device received motor error alarm during rewind due to motor encoder signal out of phase.Unable to perform the self-test, pump error alarm, pump alarm, displacement test, rewind, basic occlusion test, occlusion test, prime test and excessive no delivery test due to motor error alarms noted.
 
Event Description
Customer reported via phone call that insulin pump had multiple error alarms.Customer's blood glucose level was 230 mg/dl at the time of incident.Customer reported they were able to clear the alarm and were able to rewind the insulin pump.Customer reported that the drive support cap was normal.Advised the insulin pump will need to be replaced.Advised to discontinue use of the insulin pump and revert to backup plan.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key9340043
MDR Text Key168950001
Report Number2032227-2019-119473
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169933170
UDI-Public(01)00643169933170
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 11/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723LNAS
Device Catalogue NumberMMT-723LNAS
Device Lot NumberA6723LNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2019
Date Manufacturer Received11/14/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/27/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age77 YR
Patient Weight107
-
-