Model Number N/A |
Device Problem
Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/24/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported during an initial tka, there was hair on the articular surface upon opening the package.The surgery was completed with another device.No adverse events have been reported as a result of the malfunction.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual inspection of the returned product found a hair-like debris in the inner cavity.The packaging was returned opened.The complaint cannot be verified.The dhr was reviewed and no discrepancies relevant to the reported event were found.The likely condition of the product when leaving zimmer biomet cannot be verified.A definite root cause of the reported issue cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|