MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B2070-120
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Perforation (2001)
Date of Event 10/28/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other complaints.The reported patient effect of vessel injury, e.G.Dissection, perforation is listed in the armada 35 instructions for use as a complication which may occur as a result of percutaneous transluminal angioplasty.Based on the information provided, the balloon rupture appears to be due to case circumstances.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the 7.0x120 mm armada 35 ruptured with the first inflation at 14 atmospheres (rated burst pressure) in the heavily calcified right superficial femoral artery.In the opinion of the physician, the rupture was due to the calcified anatomy.A vessel perforation occurred which was successfully treated with a covered stent.Another same size armada 35 was used to treat the lesion.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9340984
Report Number2024168-2019-13724
Device Sequence Number2739772
Product Code LIT
UDI-Device Identifier08717648155390
UDI-Public08717648155390
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Catalogue NumberB2070-120
Device Lot Number90220G1
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/28/2019
Initial Report FDA Received Date11/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient Weight81
-
-