MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 10MM TI HINDFOOT ARTHRODESIS CANN NAIL-EX 240MM/LT-STERILE; ROD,FIXATION,INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 10MM TI HINDFOOT ARTHRODESIS CANN NAIL-EX 240MM/LT-STERILE; ROD,FIXATION,INTRAMEDULLARY Back to Search Results
Model Number 04.008.078S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932); No Code Available (3191)
Date of Event 11/05/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/ or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Investigation summary: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: manufacturing location: (b)(4), manufacturing date: 12-mar-2014, part number: 04.008.078, 10mm ti hindfoot arthrodesis cann nail ¿ ex 240mm/ left, lot number: 7630108 (non-sterile), lot quantity: (b)(4).Note: this lot number was produced as a non-sterile item.Work order traveler met all inspection acceptance criteria.Inspection sheet, in-process acceptance sheet, op038354 rev a met all inspection acceptance criteria.Inspection sheet, ns037189 rev c met all inspection acceptance criteria.Inspection sheet, final inspection, 04fi008010 rev d met all inspection acceptance criteria.Packaging label log lppf, lmd/lpf rev ab was reviewed and determined to be conforming.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device history review: 14-nov-2019: dhr reviewed by: (b)(4).This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2019, the patient underwent a revision surgery due to chronic osteomyelitis, infection, inflammation, nonunion of joint, and delayed healing.One (1) hindfoot nail, one (1) spiral blade, one (1) 6mm unknown locking screw, and one (1) 5mm unknown locking screw were removed.All hardware was intact.After the removal, the extremity of the patient was i&d using ria.There was no surgical delay.  it is unknown if procedure was successfully completed.Patient status was good.This complaint involves four (4) devices.This is 1 of 4 for report (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10MM TI HINDFOOT ARTHRODESIS CANN NAIL-EX 240MM/LT-STERILE
Common Device Name
ROD,FIXATION,INTRAMEDULLARY
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key9342513
Report Number2939274-2019-62275
Device Sequence Number16847060
Product Code HSB
UDI-Device Identifier10886982087218
UDI-Public(01)10886982087218
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K051678
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.008.078S
Device Catalogue Number04.008.078S
Device Lot Number7630108
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/04/2019
Initial Report FDA Received Date11/19/2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
Patient Weight93
-
-