| Model Number |
657318S |
| Medical Device Problem Codes |
Material Erosion (1214); Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Variax clavicle operation.When the locking screw was inserted, a thread-like metal piece came out from between the plate and the screw.Replaced with a new locking screw of the same size.The operation ended successfully.
|
| |
|
Event or Problem Description
|
|
Variax clavicle operation.When the locking screw was inserted, a thread-like metal piece came out from between the plate and the screw.Replaced with a new locking screw of the same size.The operation ended successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed.The inspection revealed that the foreign material is a metallic wire.The microscope investigation showed that the threaded part of the screw is missing some material.Moreover, the metallic wire has the same color as the screw.Therefore, most probably, while tightening the screw, the metallic wire came out.The wire came out most likely because the screw was inserted with too big of an angle (the smart lock technology allows the surgeon to aim and lock the screw within a 30° cone), which could have caused the friction between the screw and a sharp part of the plate (the titanium lip for instance), leading to the removal of a part of the material.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to be user related.If any further information is provided, the investigation report will be updated.
|
| |
|