| Model Number |
G34785 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pma/510(k) #: k142688.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
Dr.(b)(6) got some sample in the first, second penetration but on the third 1.5cm of needle was broken.
|
| |
|
Additional Manufacturer Narrative
|
|
Pma/510(k) #k142688.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
Dr.(b)(6) got some sample in the first, second penetration but on the third 1.5cm of needle was broken.
|
| |
|
Event or Problem Description
|
|
Dr.(b)(6) got some sample in the first, second penetration but on the third 1.5cm of needle was broken.
|
| |
|
Additional Manufacturer Narrative
|
|
Pma/510(k) #k142688.The echo-hd-3-20-c device of lot number c1627561 involved in this complaint was returned for evaluation, with the original packaging.The packaging was open on receipt.With the information provided, a physical examination and document based investigation was conducted.The device related to this occurrence underwent a laboratory evaluation on the 13-nov-2019.Needle broken at distal end.The broken part of the needle was not returned.Prior to distribution, all echo-hd-3-20-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the relevant manufacturing records (c1627561) revealed no discrepancies that could have contributed to this complaint.Upon review of complaints, this failure mode has not occurred previously with this lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1627561.There is no evidence to suggest that the customer did not follow the instructions for use."visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use." a definitive root cause could not be determined as the circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to hard lesion.Complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|