COOK IRELAND LTD ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY Back to Search Results
Model Number G34785
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/23/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) #: k142688.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
Dr.(b)(6) got some sample in the first, second penetration but on the third 1.5cm of needle was broken.
 
Additional Manufacturer Narrative
Pma/510(k) #k142688.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
Dr.(b)(6) got some sample in the first, second penetration but on the third 1.5cm of needle was broken.
 
Event or Problem Description
Dr.(b)(6) got some sample in the first, second penetration but on the third 1.5cm of needle was broken.
 
Additional Manufacturer Narrative
Pma/510(k) #k142688.The echo-hd-3-20-c device of lot number c1627561 involved in this complaint was returned for evaluation, with the original packaging.The packaging was open on receipt.With the information provided, a physical examination and document based investigation was conducted.The device related to this occurrence underwent a laboratory evaluation on the 13-nov-2019.Needle broken at distal end.The broken part of the needle was not returned.Prior to distribution, all echo-hd-3-20-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the relevant manufacturing records (c1627561) revealed no discrepancies that could have contributed to this complaint.Upon review of complaints, this failure mode has not occurred previously with this lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1627561.There is no evidence to suggest that the customer did not follow the instructions for use."visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use." a definitive root cause could not be determined as the circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to hard lesion.Complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE
Common Device Name
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key9346433
Report Number3001845648-2019-00614
Device Sequence Number11318658
Product Code FCG
UDI-Device Identifier00827002347854
UDI-Public(01)00827002347854(17)220703(10)C1627561
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/03/2022
Device Model NumberG34785
Device Catalogue NumberECHO-HD-3-20-C
Device Lot NumberC1627561
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/12/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/23/2019
Event Location Hospital
Initial Date Received by Manufacturer 10/23/2019
Supplement Date Received by Manufacturer10/23/2019
11/23/2019
Initial Report FDA Received Date11/20/2019
Supplement Report FDA Received Date12/19/2019
02/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-