MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRITANIUM BPLATE TRIATHLON S6; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRITANIUM BPLATE TRIATHLON S6; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5536B600
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Injury (2348); Inadequate Osseointegration (2646)
Date of Event 10/23/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.Device evaluated by manufacturer? not returned to the manufacturer.
 
Event or Problem Description
As reported: ".Patient is 3 year post primary left tka.Doctor told me that the patient has had pain and recurrent effusions and his imaging studies suggest a loose tibial component.The patient was scheduled for surgery on (b)(6) 2019.Doctor used an osteotome circumferentially around the tibial component to loosen the tibial baseplate [revision operative report states "i was able to place an osteotome beneath the component between the bone and implant.This was a press-fit implant originally.Was able to get the osteotome underneath the component nearly all the way around and then it easily lifted up.It was clearly not well fixed.There was no bone loss during the removal."].He inspected the other implants and said after surgery they appeared to be well fixed.He did a tibial clean up cut and prepped for a universal cemented baseplate and 12x50 stem.He trialed a 16mm ps poly trial and asked for the actual implant.There will be no explants because they were discarded by the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRITANIUM BPLATE TRIATHLON S6
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9347931
Report Number0002249697-2019-03829
Device Sequence Number16890407
Product Code MBH
UDI-Device Identifier07613327041514
UDI-Public07613327041514
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2021
Device Catalogue Number5536B600
Device Lot NumberCTD10601
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/23/2019
Initial Report FDA Received Date11/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age71 YR
-
-