DEPUY IRELAND - 9616671 ATTUNE FB TIB BASE SZ 9 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 150600009
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Pain (1994); No Code Available (3191)
Date of Event 06/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2014, the patient underwent total left knee arthroplasty due to osteoarthritis.The patella was resurfaced.The surgeon reported no intraoperative complications.The patient was implanted with the attune knee system and smartset bone cement x 2.On (b)(6) 2018, the patient underwent a left knee revision due to loosening of the tibial component, instability and pain.The surgeon reported the patient was grossly unstable laterally.He reported the tibial component was loose needing one strike of the mallet which resulted the component to de-bond 100 percent from the underlying cement.He noted a well-fixed femoral component.The surgeon reported the patella had some wear and so was revised.The patient was implanted with a competitor system and there were no complications to the procedure.Doi: (b)(6) 2014.Dor: (b)(6) 2018.(rt knee).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB TIB BASE SZ 9 CEM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9348072
Report Number1818910-2019-117163
Device Sequence Number13992568
Product Code JWH
UDI-Device Identifier10603295042433
UDI-Public10603295042433
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Catalogue Number150600009
Device Lot Number7881152
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/24/2019
Supplement Date Received by Manufacturer12/13/2019
Initial Report FDA Received Date11/20/2019
Supplement Report FDA Received Date12/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
-
-