| Model Number |
N/A |
| Medical Device Problem Codes |
Positioning Failure (1158); Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/07/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 0001822565-2019-04925.Concomitant medical products: unknown nexgen tibial tray.Report source: (b)(6).Customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon was unable to clip the articular surface into the tibia plate.There is no additional information at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up is being submitted to relay additional information.The reported event is not confirmed.Product was returned and visual examination of the returned product identified that the dovetail feature exhibits damage (flared out).Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no related deviations anomalies identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to update associated medical product.Medical products: stemmed nonaugmentable tibial component option cr,ps,lps item# (b)(4).Lot# 64500603.
|
| |
|