As the device was received in a condition was contradictory to the complaint description.As received, the pushwire was found detached at the distal hypotube.The re-sheathing pad was found separate from the pushwire.This condition was not reported initially.In addition, pipeline flex pushwire was inadvertently cut during analysis.The tip coil was found bent.The distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.The re-sheathing marker and proximal bumper were not found.The distal hypotube and ptfe shrink tubing were found to be intact with no signs of elongation.No bends were found on the pushwire.The ends of the detached pushwire were sent out for sem (scanning electron micrographic) / eds (energy dispersive spectroscopy) elemental analysis.Based on the device analysis, the distal wire possibly detached due to tensile failure.However, the cause of the event could not be determined.If information is provided in the future, a supplemental report will be issued.
|