• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. 3D KNEE INSERT, 3D EX SZ 8LT 9MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. 3D KNEE INSERT, 3D EX SZ 8LT 9MM Back to Search Results
Model Number 391-09-708
Device Problem Unstable (1667)
Patient Problem Failure of Implant (1924)
Event Date 10/24/2019
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was reported as instability. The previous surgery and the surgery detailed in this event occurred 1 year and 7 months apart. The healthcare professional indicated there was no delay in surgery and another suitable device was available for use. The revision surgery was completed as intended. The device was disposed of at hospital and not made available to djo surgical for examination. A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements. There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event. The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery. Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review. The root cause of this complaint was a revision surgery due to instability. There were no findings during this evaluation that indicate the reported device was defective. No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event. There are multiple factors that may contribute to the event that are outside the control of djo surgical such as inadequate soft tissue support, excessive range of motion, poor bone density, patient bone deterioration, patient activities or trauma. There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event Description
Revision surgery - the patient knee had become unstable, a larger insert was implanted to stabilize the knee.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name3D KNEE
Type of DeviceINSERT, 3D EX SZ 8LT 9MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key9349524
MDR Text Key167342855
Report Number1644408-2019-01116
Device Sequence Number1
Product Code OIY
UDI-Device Identifier00888912118989
UDI-Public(01)00888912118989
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091956
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model Number391-09-708
Device Catalogue Number391-09-708
Device Lot Number285G1949
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/24/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 11/20/2019 Patient Sequence Number: 1
-
-