STRYKER GMBH DISTAL LATERAL FIBULA PLATE, 7 HOLE, STERILE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Catalog Number 40-20907S
Medical Device Problem Code Difficult or Delayed Positioning (1157)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/23/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device remains implanted.
 
Event or Problem Description
During variax fibula surgery, a locking screw did not lock to the most distal hole of the plate.The surgeon inserted the screw as it is and finished the surgery.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidence was provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
During variax fibula surgery, a locking screw did not lock to the most distal hole of the plate.The surgeon inserted the screw as it is and finished the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISTAL LATERAL FIBULA PLATE, 7 HOLE, STERILE
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key9349741
Report Number0008031020-2019-01869
Device Sequence Number14041409
Product Code KTT
UDI-Device Identifier04546540604651
UDI-Public04546540604651
Combination Product (Y/N)N
PMA/510(K) Number
K081284
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2023
Device Catalogue Number40-20907S
Device Lot Number1000303248
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/23/2019
Supplement Date Received by Manufacturer12/19/2019
Initial Report FDA Received Date11/20/2019
Supplement Report FDA Received Date01/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-