(b)(4).Unable to perform the displacement test and the p-cap/reservoir will not lock properly due to missing retainer.Device received with missing reservoir tube o-ring, broken reservoir tube lip and fading serial number label.The power management tool confirmed power error detected, low battery alert, power loss alarm, triggered when the backup battery loaded voltage (loaded vlith) was less than 3.5v for 4 consecutive hours due to connector resistance pcb 1.After disconnecting and reconnecting the internal battery connector on pcb 1, the pump was monitored and functioned properly.
|
This report is part of a retrospective review and remediation efforts in response to a warning letter.Updated h9: z-0955-2020.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|