| Catalog Number |
5532-G-209 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/24/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Pmh reported that the locking wire was loosened when they implanted into the baseplate.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding a seating/locking issue involving a triathlon insert was reported.The event was confirmed based on evaluation of returned device.Method & results: product evaluation and results:damage consistent with the attempted implantation was observed on the anterior portion of the distal surface.No identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Clinician review: not performed as no medical records were provided for review.Product history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: the event was confirmed based on the evaluation of the returned device, however, the root cause could not be determined because insufficient information was provided.Additional information including photo/video of the alleged event are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.Product surveillance will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
Pmh reported that the locking wire was loosened when they implanted into the baseplate.
|
| |
|