• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5532-G-209
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Pmh reported that the locking wire was loosened when they implanted into the baseplate.
 
Additional Manufacturer Narrative
An event regarding a seating/locking issue involving a triathlon insert was reported.The event was confirmed based on evaluation of returned device.Method & results: product evaluation and results:damage consistent with the attempted implantation was observed on the anterior portion of the distal surface.No identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Clinician review: not performed as no medical records were provided for review.Product history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: the event was confirmed based on the evaluation of the returned device, however, the root cause could not be determined because insufficient information was provided.Additional information including photo/video of the alleged event are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.Product surveillance will continue to monitor for trends.
 
Event or Problem Description
Pmh reported that the locking wire was loosened when they implanted into the baseplate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9353249
Report Number0002249697-2019-03836
Device Sequence Number10858182
Product Code MBH
UDI-Device Identifier07613327046205
UDI-Public07613327046205
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Catalogue Number5532-G-209
Device Lot NumberRN7A3T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2019
Initial Date Received by Manufacturer 10/24/2019
Supplement Date Received by Manufacturer12/16/2019
Initial Report FDA Received Date11/21/2019
Supplement Report FDA Received Date01/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-