Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was contrast and blood in the inflation lumen and balloon, and blood in the guidewire lumen.The balloon was loosely folded.The device was soaked in a water bath for two days to loosen the blood and contrast in the device.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There was a pinhole at the distal marker band.There was no marker band damage detected.The device was functionally tested with a.014" guidewire.No issues were detected.Inspection of the remainder of the device presented no other damage or irregularities.
|