• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTFLOW, INC. FFRCT; HEARTFLOW FFRCT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTFLOW, INC. FFRCT; HEARTFLOW FFRCT Back to Search Results
Model Number FFRCT 2.45.0.1
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Information (3190)
Date of Event 10/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As part of heartflow's quality monitoring process, we identified a potential false negative.(b)(4).The investigation identified a potential false negative in the lcx region; after the initial analysis was delivered, a second analysis completed as part of ongoing quality review resulted in a decrease in the ffrct value from 0.75 to <0.50 on the distal lcx.While heartflow has identified this issue, the physician has not provided feedback, nor indicated a safety event with the patient.
 
Event or Problem Description
Heartflow identified a potential false negative result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FFRCT
Common Device Name
HEARTFLOW FFRCT
Manufacturer (Section D)
HEARTFLOW, INC.
1400 seaport blvd., bldg. b
redwood city CA 94063 5594
MDR Report Key9354852
Report Number3011276938-2019-00050
Device Sequence Number2055414
Product Code PJA
UDI-Device Identifier00853341006015
UDI-Public(01)00853341006015(10)FFRCT_2.45.0.1(11)2019-10-28
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K161772
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFFRCT 2.45.0.1
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2019
Initial Report FDA Received Date11/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-