• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GEN II CEMENTED TIBIAL B/PLATE SZ2 RIGHT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GEN II CEMENTED TIBIAL B/PLATE SZ2 RIGHT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 71420180
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 10/18/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that revision surgery was performed on (b)(6) 2019 due to the infection and loosening.Primary surgery was performed on (b)(6) 2011 and gii fem component(71420052), gii tib base (71420180), legion insert (71453202), patella (71420580).All of the implanted devices were revised.The revised devices will not be returned.The doctor does not consider as the failure of device.
 
Additional Manufacturer Narrative
The devices, used in treatment, were not returned for evaluation and the reported event could not be confirmed.The clinical / medical investigation concluded that it has been communicated that no further information would be forthcoming.Without supporting clinical/medical documents a thorough investigation cannot be performed.Should information become available this complaint can be re-assessed.A review of complaint history did not reveal additional complaints for the listed batches.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the sterilization records revealed the batch was sterilized within normal parameters.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.Infection is a potential complication associated with any surgery.Some potential probable causes could include contamination, patient reaction or a post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaints will be reopened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GEN II CEMENTED TIBIAL B/PLATE SZ2 RIGHT
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9356854
Report Number1020279-2019-04127
Device Sequence Number2220083
Product Code JWH
UDI-Device Identifier03596010206329
UDI-Public03596010206329
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71420180
Device Catalogue Number71420180
Device Lot Number10GT44466
Initial Date Received by Manufacturer 10/23/2019
Supplement Date Received by Manufacturer10/23/2019
Initial Report FDA Received Date11/21/2019
Supplement Report FDA Received Date03/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-