Model Number 1217-30-052
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); No Code Available (3191)
Date of Event 10/28/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #(b)(4).(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The primary surgery was performed on (b)(6) 2018 via tha.It was reported that removal surgery was performed on (b)(6) 2019 due to infection.The stem (p/n: 3l92502), the head (p/n: 136550000), the liner (p/n: 122436052), the cup 8p/n: 121730052) were explanted and indwelling the cement beads.The surgery was competed, and it was unknown whether there was a surgical delay or not.No further information is available.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MULTIHOLE W/GRIPTION 52MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9357226
Report Number1818910-2019-117501
Device Sequence Number11323440
Product Code KWA
UDI-Device Identifier10603295009986
UDI-Public10603295009986
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1217-30-052
Device Catalogue Number121730052
Device Lot NumberHZ2214
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/27/2019
Supplement Date Received by Manufacturer01/15/2020
Initial Report FDA Received Date11/21/2019
Supplement Report FDA Received Date01/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-