• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MITG - OKLAHOMA CITY TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MITG - OKLAHOMA CITY TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 72204064
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/05/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the base of the blade was detached.There was no patient involvement.
 
Additional Manufacturer Narrative
This event has been reassessed and found to be a non-mdr reportable complaint.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Common Device Name
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
MITG - OKLAHOMA CITY
75 s. meridian ave
oklahoma city OK 73107
MDR Report Key9358852
Report Number1643264-2019-20027
Device Sequence Number2219447
Product Code HIH
UDI-Device Identifier10884521744097
UDI-Public10884521744097
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K161763
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/22/2020
Device Model Number72204064
Device Catalogue Number72204064
Device Lot Number4732878
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/05/2019
Supplement Date Received by Manufacturer11/18/2019
Initial Report FDA Received Date11/21/2019
Supplement Report FDA Received Date12/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-