| Model Number |
10602 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is combination product.Date of event is estimated as (b)(6) 2019 based on the date the malfunction form was completed.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.The target lesion was around a sharp (severely tortuous) and severely calcified turn in the mid circumflex artery.Three 3.00 x 16 mm synergy drug eluting stents were introduced but each one got caught on the calcium resulting in stent deformation.There was no patient harm and the patient was referred to a larger facility for atherectomy.
|
| |
|