| Model Number |
MMT-723RNAP |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 290 mg/dl at the time of the incident.Customer was assisted with troubleshooting.The customer also reported that they were able to clear and not able to rewind the insulin pump.The customer was advised that the insulin pump needed to be replaced and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Search Alerts/Recalls
|