MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number RONYX35018UX
Medical Device Problem Codes Fracture (1260); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/13/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device returned with a detachment on the hypotube 48.5cm distal to the strain relief.The hypotube material was oval and jagged on both sides of the detachment site.A kink was evident on the hypotube 4.5cm proximal to the detachment site.No other damage evident to the remainder of the device.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During a procedure an attempt was made to use a resolute onyx rx coronary drug eluting stent to treat a lesion in the right coronary artery (rca).The device was inspected with no issues noted.Negative prep was performed without issue.The device was not kinked and re-straightened during use.It was reported that the stent shaft fractured while advancing the device through the guide catheter.It was able to be removed by removing the catheter.The patient was reported to be alive with no injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key9361240
Report Number9612164-2019-04852
Device Sequence Number3232005
Product Code NIQ
UDI-Device Identifier00643169557086
UDI-Public00643169557086
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/22/2021
Device Model NumberRONYX35018UX
Device Catalogue NumberRONYX35018UX
Device Lot Number0009837501
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2019
Initial Date Received by Manufacturer 11/18/2019
Supplement Date Received by Manufacturer01/13/2020
Initial Report FDA Received Date11/22/2019
Supplement Report FDA Received Date02/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
-
-