DEPUY MITEK LLC US MILAGRO INTSCR 7X23MM *EA; ORTHOPAEDIC BONE SCREW, BIOABSORBABLE Back to Search Results
Model Number 231800
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/05/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.Udi: (b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
This is report 1 of 2 for the same event.It was reported by the affiliate in (b)(6) that during the procedure of an ankle ligament reconstruction, it was observed that the 6x12 milagro interfrence screw was broken off into two pieces upon screwing.It was reported that the pieces were removed from the patient.It was reported that the same thing happened when screwing the milagro intscr 7x23mm *ea which was broken off into two pieces.It was reported that one of the pieces was not able to be removed because it was insert into the bone.The procedure was completed using another device.No patient consequence and no surgical delay was reported.There was no delay in the surgical procedure.There was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
Product complaint #
=
> pc(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary
=
> the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number (2l06392), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
ORTHOPAEDIC BONE SCREW, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9363121
Report Number1221934-2019-59711
Device Sequence Number14196529
Product Code MAI
UDI-Device Identifier10886705010684
UDI-Public10886705010684
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Model Number231800
Device Catalogue Number231800
Device Lot Number2L06392
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/13/2019
Supplement Date Received by Manufacturer01/22/2020
Initial Report FDA Received Date11/22/2019
Supplement Report FDA Received Date01/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-