|
Product complaint # (b)(4).Additional information: concomitant medical products, device manufacture date.Evaluation: one empty labeled winding former and two needle-suture piece of product code hs6822, lot mph335 were received for analysis.The product code is double armed.During the visual inspection of the opened sample (two needle/suture pieces), the swage and attachment area were noted to be as expected.However, marks on the body needle and the end of the suture pieces cut that appears to be by the use of a surgical instrument could be observed.The other section of the suture was not returned for analysis and due to the condition of the sample the functional test can¿t be performed.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Due to condition of the opened sample, the assignable cause of performance breakage suture suggests an improper handling and the complaint one empty labeled winding former and two needle-suture piece of product code hs6822, lot mph335 were received for analysis.The product code is double armed.During the visual inspection of the opened sample (two needle/suture pieces), the swage and attachment area were noted to be as expected.However, marks on the body needle and the end of the suture pieces cut that appears to be by the use of a surgical instrument could be observed.The other section of the suture was not returned for analysis and due to the condition of the sample the functional test can¿t be performed.Due to condition of the opened sample, the assignable cause of performance breakage suture suggests an improper handling.
|