| Catalog Number |
0067010 |
| Medical Device Problem Codes |
Break (1069); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject product was received for evaluation.Visual evaluation found the battery pack was within the handle and the battery latch was noted to be broken off.The battery pack was received firmly within the device.A broken battery latch is not indicative of a manufactured condition and based on the use facility reporting that the batteries were changed during use, the damage battery latch most likely occurred due to customer handling.A review of the manufacturing records was performed and found that the lot was manufactured to specification.
|
| |
|
Event or Problem Description
|
|
It was reported per the user facility that during use of the simpulse varicare, the device was leaking fluid at the connection point of where the tip attaches to the handle.It was reported that there was no damage to the device.It was reported that the device was only used with one patient and the batteries were changed during the time of use, towards the end of treatment.There was no patient injury reported.The subject product was returned and during visual inspection it was noted the battery case latch tab was broken and not returned with the product.
|
| |
|