| Model Number |
MODEL 3000 |
| Medical Device Problem Codes |
Computer Software Problem (1112); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported via a programming history database periodic review that the patient had high impedance.The high impedance was observed with v1.5 software.For model 102/102r generators programmed to output currents > 1 ma , it was observed that system diagnostics using m3000 v1.0 and m250 software would display ok lead impedance while system diagnostics using m3000 v1.5 software would display a false high impedance.No other relevant information has been received to date.
|
| |
|