Catalog Number UNK FIXATION PIN |
Device Problems
Break (1069); Mechanical Jam (2983); Physical Resistance/Sticking (4012)
|
Patient Problem
Not Applicable (3189)
|
Event Date 10/29/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the a headed pin got stuck in a tc3 cut block and when trying to remove it, the head of the pin was broken off and became stuck in the pin driver.A surgical delay of 5 minutes.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: product complaint # (b)(4).Investigation summary examination of the returned device confirmed the reported event.The investigation attributed the root cause to unintended misuse/misapplication of the device and did not indicate that that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot null.Device history batch null.Device history review null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: product complaint # (b)(4).Investigation summary examination of the returned device confirmed the reported event.The investigation attributed the root cause to unintended misuse/misapplication of the device and did not indicate that that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot null.Device history batch null.Device history review null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|