• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION SPECTRANETICS TURBO ELITE LASER ATHERECTOMY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION SPECTRANETICS TURBO ELITE LASER ATHERECTOMY CATHETER Back to Search Results
Model Number 420-006
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Codes Burning Sensation (2146); Radiation Exposure, Unintended (3164)
Date of Event 11/07/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The physician in this procedure was male.Patient information not obtained for this event; there is no reported patient harm.
 
Event or Problem Description
A peripheral vascular intervention commenced to treat a lesion in the patient's superficial femoral artery (sfa).During use of the spectranetics turbo elite device, it was reported that the physician felt a sensation which felt warm through his glove.He expressed discomfort in his finger.The location of this occurrence was reportedly where the physician's fingers were placed while the device was in use, outside of the patient's body.It was believed that the physician's glove remained intact, and it was reported that no first aid or intervention was required for this area on the physician's finger.The procedure was completed with the device and the patient survived the procedure.In addition, due to the description of the event reported to the manufacturer, this event is being reported due to the potential for exposure to manufacturing materials as well as inadvertent exposure to laser energy/radiation, in the event that the device was broken.The manufacturer is anticipating the return of the device for evaluation but has not yet arrived at the manufacturer.
 
Additional Manufacturer Narrative
Device was returned to manufacturer on 02 dec 2019.Device evaluation performed.Patient code 3164: in the initial mdr, code 3164 was used.After device evaluation, it was confirmed that there was no potential for exposure to manufacturing materials nor unintended radiation exposure.Code no longer appropriate in this event.Method, results and conclusions codes now populated based on results of device evaluation.Device evaluation: the device evaluation was performed and completed by a cross functional engineering team on 05 dec 2019.Visual inspection of the device confirmed no damage to the catheter.The catheter was plugged into simulated laser light; no breach of the device jacket was seen.Under the microscope, white residue is seen on the catheter at 78cm (unknown substance, however this residue would not cause harm to the physician).Again confirmed there was no breach to the jacket.The evaluation could not determine the cause of the experienced failure.Due to no malfunction being present for this turbo elite device, this is no longer a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS TURBO ELITE LASER ATHERECTOMY CATHETER
Common Device Name
TURBO ELITE
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
MDR Report Key9366094
Report Number1721279-2019-00207
Device Sequence Number11506595
Product Code MCW
UDI-Device Identifier00813132024758
UDI-Public(01)00813132024758(17)210718(10)FBA19G16A
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K170059
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/07/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/18/2021
Device Model Number420-006
Device Catalogue Number420-006
Device Lot NumberFBA19G16A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2019
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/07/2019
Supplement Date Received by Manufacturer12/05/2019
Initial Report FDA Received Date11/22/2019
Supplement Report FDA Received Date12/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-