DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22719-BLU
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/10/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that signal loss over one hour occurred.The product was evaluated.An external visual inspection was performed and passed.A charge and boot test was performed and passed.A pairing test was performed and passed.A receiver functional test was performed and passed.There was no data within the investigation window.The allegation was undetermined.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
MDR Report Key9366882
Report Number3004753838-2019-093894
Device Sequence Number11370577
Product Code PQF
UDI-Device Identifier30386270000270
UDI-Public30386270000270
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22719-BLU
Device Catalogue NumberSTK-GF-BLU
Device Lot Number5253436
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2019
Initial Report FDA Received Date11/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
-
-