| Catalog Number |
9569313 |
| Medical Device Problem Codes |
Break (1069); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
10/25/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Other: device fragment in patient.Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent oblique lumbar interbody fusion at l3-l4-l5 due to lumbar canal stenosis and spondylolisthesis at these levels.The patient had a very hard bone.Intra-op, the tip of the pin broke when the alleged pin was being inserted into l4 vertebral body.Another pin was used to complete the procedure successfully.The broken part was tried to be removed but could not be removed.Hence, wound closure was performed, with the fragment of the broken tip remaining in the vertebral body.There was a delay of less than 60 minutes in overall procedure time as a result of this event.No patient complications have been reported yet.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: product analysis: visually confirmed tip broken off the instrument.The broken tip was not returned for analysis.Severe damage noted on fracture surface with multiple areas of secondary damage, consistent with multiple impacts after fracture.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|