| Catalog Number |
07K35-68 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
|
| |
|
Event or Problem Description
|
|
The customer reported an increase in false repeat reactive prism chagas results on multiple donors that did not confirm by esa chagas confirmatory testing.No impact to donor management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of complaint data for the product and likely cause abbott prism chagas lot 05069m700 identified normal complaint activity.No customer returns were available for evaluation.The performance of the likely cause lot was investigated by completing a review for non-conformances, potential non-conformances and deviations related to the likely cause lot.This review did not identify any non-conformances, potential non-conformances or deviations.A review of overall customer field data was performed regarding initial reactive rate (irr) and repeat reactive rate (rrr) for lot 05069m700.The irr and the rrr of the lot is within package insert specifications.A review of labeling concluded that the issue is sufficiently addressed.No product deficiency was identified.
|
| |
|