• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL COMBO BEATH PIN EA; CANNULATED SURGICAL DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL COMBO BEATH PIN EA; CANNULATED SURGICAL DRILL Back to Search Results
Model Number 232453
Medical Device Problem Codes Break (1069); Material Rupture (1546)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 11/07/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/ or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported by the affiliate in (b)(6) that the wire guide of the combo beath pin ea device was ruptured.The procedure was completed using another wire with a 20 minute delay.There was a delay in the surgical procedure.There was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> according to the information provided, it was reported a rupture of the wire guide during the procedure.The complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.However photos were provided.Upon visual inspection of the photos, the device was observed broken and fully separated.The reported complaint was confirmed.The photos does not provide enough evidence to determine root cause.A possible root cause can be associated to an excessive force applied during use, however it cannot be conclusively affirmed.Hands on analysis should provide the evidence necessary to confirm the root cause.A manufacturing record evaluation was performed for the finished device [3l39743] number, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot
=
> a manufacturing record evaluation was performed for the finished device [3l39743] number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMBO BEATH PIN EA
Common Device Name
CANNULATED SURGICAL DRILL
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9370793
Report Number1221934-2019-59763
Device Sequence Number2902384
Product Code HWE
UDI-Device Identifier10886705024544
UDI-Public10886705024544
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model Number232453
Device Catalogue Number232453
Device Lot Number3L39743
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2019
Initial Date Received by Manufacturer 11/21/2019
Supplement Date Received by Manufacturer11/21/2019
01/27/2020
Initial Report FDA Received Date11/25/2019
Supplement Report FDA Received Date12/20/2019
01/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-