| Model Number |
00884450200018 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/28/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a trans jugular liver biopsy procedure, the catheter tip detached within the patient's hepatic vein.The physician acquired antegrade venous access and during catheter manipulations the tip detached.The physician successfully removed the catheter tip with a vascular snare device with no additional consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The complaint could not be confirmed and a root cause could not be determined.A review of the device history and complaint database could not be performed since the lot number was not provided.
|
| |
|