• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL Back to Search Results
Catalog Number PWF030F
Health Effect - Clinical Codes Itching Sensation (1943); Burning Sensation (2146); Reaction (2414); Patient Problem/Medical Problem (2688)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the patient stated she had terrible reaction to the product; took her a month to recover from the issue.Per additional information received from the customer, she experienced burning and itching.She used ketamine cream and lidocaine.In addition, she had a vaginal suppository.
 
Event or Problem Description
It was reported that the patient stated she had terrible reaction to the product; took her a month to recover from the issue.Per additional information received from the customer, she experienced burning and itching.She used ketamine cream and lidocaine.In addition, she had a vaginal suppository.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿contraindications: -patients with urinary retention warnings: -do not use purewick female external catheter with bedpan or any material that does not allow for sufficient airflow.-to avoid potential skin injury, never push or pull the purewick female external catheter against the skin during placement or removal, -never insert the purewick female external catheter into vagina, and anal canal or other body cavities.-discontinue use if an allergic reaction occurs.-after use, this product may be a potential biohazard.Dispose of in accordance with applicable local, state and federal laws and regulations.Precautions: -not recommended for patients who are: -agitated, combative or unconperative and might removw the purewick female external catheter.-having frequent episodes of bowel incontinence without a fecal management system in place.-experiencing skin irritation or breakdown at the site -experiencing moderate/heavy menstruation and cannot use a tampon.-do not use barrier cream on the perineum when using the purewick female external catheter.Barrier cream may impede suction -not recommended for use on patients who have undergone recent surgery of the external urogenital tract.-always assess skin for compromise nd perform perineal care prior to placement of a new purewick female external catheter.-maintain suction until the purewick female external catheter is fully removed from he patient to avoid urine backflow.Recommendations: -perform each step with clean technique.N the home setting, wash hands thoroughly before device placement.-prior to connecting the purewick female external catheter to hospital wall suction tubing, verify suction function by covering the open end of the suction tubing with one hand and observing the pressure dial.If the pressure does not increase when the line is covered, verify that the tubing is secured, connected and not kinked.-ensure the purewick female external catheter remains in the correct position after turning the patient.Remove the purewick female external catheter prior to ambulation.-properly placing the purewick female external catheter snugly between the labia and gluteus holds the purewick female external catheter in place for most patients.Mesh underwear may be useful for securing the purewick female external catheter for some patients.-assess device placement and patient's skin at least every 2 hours.-replace the purewick female external catheter every 8-12 hours or when soiled with feces or blood.-change suction tubing per hospital protocol or at least every thirty (30) days." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUREWICK FEMALE EXTERNAL
Common Device Name
PUREWICK
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9371564
Report Number1018233-2019-07527
Device Sequence Number3103645
Product Code NZU
UDI-Device Identifier00801741163081
UDI-Public(01)00801741163081
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberPWF030F
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/06/2019
Supplement Date Received by Manufacturer01/01/2020
Initial Report FDA Received Date11/25/2019
Supplement Report FDA Received Date01/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-