| Catalog Number |
SMXP130RH |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: unknown.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) female patient underwent a breast surgery with artoura breast tissue expander 475cc and experienced unspecified complications resulting in removal.The patient alleged they had a defective device.As a result, the patient underwent explantation on (b)(6) 2019.
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2019 mentor completed evaluation of the device.Device evaluation summary: during visual inspection of the device it was observed with no apparent damage.Leak testing was performed in accordance with mentor procedures and no leak sites were detected.No anomalies were observed.It should be noted that as part of our quality process, each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Manufacturer¿s reference number: (b)(4).
|
| |
|