| Model Number |
MMT-723LNAH |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer's family member reported via phone call that the insulin pump completely shut down.The customer¿s blood glucose level was unknown at the time of the incident.Customer stated the contacts on the battery cap were not missing or damaged.Customer stated battery compartment was neither damaged nor corroded.Customer stated that the spring was neither damaged nor corroded.The customer was assisted with troubleshooting and display did not returned back.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device power up properly after battery installation.No blank display noted.Device was received with normal operating currents and passed self-test, a21 error test and displacement test.No unexpected off no power, bad battery, low battery, weak battery, battery out limit or failed battery test alarms noted during testing.
|
| |
|