DEPUY ORTHOPAEDICS INC US SELF CENT HIP 40X28 RUST; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI Back to Search Results
Model Number 1035-40-000
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Pain (1994); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
Date of Event 10/30/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Surgeon had a patient with a 40 mm bipolar head and 40 mm +4 10 degree liner in situ.He had to remove the self centering bipolar head because the poly insert in the head¿s rim that clips into big metal head, broke.Patient status/outcome/consequences: yes.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. .
 
Event or Problem Description
Additional information received that the patient experienced pain and immobility.There was no surgical delay and when patient presented with pain and immobility, surgeon had x-rays done and saw the position of the femoral head implant was altered and decided to have it revised.The affected side was the right hip and the date of implantation was 2018.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: product complaint (b)(4).Investigation summary: the device was reviewed by bioengineering and a report was received stating it is not possible to determine if a potential product issue was present root cause undetermined.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF CENT HIP 40X28 RUST
Common Device Name
HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9375692
Report Number1818910-2019-118013
Device Sequence Number11535903
Product Code KWY
UDI-Device Identifier10603295003304
UDI-Public10603295003304
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K812672
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1035-40-000
Device Catalogue Number103540000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2019
Initial Date Received by Manufacturer 10/31/2019
Supplement Date Received by Manufacturer11/27/2019
01/14/2020
01/16/2020
Initial Report FDA Received Date11/26/2019
Supplement Report FDA Received Date12/26/2019
01/27/2020
01/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICUL/EZE BALL 28 +1.5 GR; ARTICUL/EZE BALL 28 +1.5 GR
Outcome Attributed to Adverse Event Required Intervention;
-
-