| Model Number |
8900 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: used (b)(6) 2019 as event date since the correct date was not provided.
|
| |
|
Event or Problem Description
|
|
It was reported that a wire break occurred.Vascular access was obtained via the radial artery.The 70% stenosed lesion was located in the non tortuous and calcified, 20mm target lesion.This 185cm comet pressure guidewire was selected along with a opticross 6 imaging catheter which was loaded onto the comet wire and were delivered beyond the lesion.After imaging was performed and the opticross imaging catheter was withdrawn and the physician felt a "snap".Angiogram revealed that the comet wire had separated into two pieces in the proximal segment of the radiopaque tip.The physician retracted the comet wire and imaging catheter into the guide catheter as one unit.All parts of the wire were retrieved from the body.Another of the same device was used to completed the procedure.No patient complications were reported.
|
| |
|