| Catalog Number |
72202961S |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a shoulder arthroscopy the scope lens did not attach properly.The procedure was completed with a back up device with a delay greater than 30 minutes.No patient injuries were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional narrative.The device, which was used in a procedure, was returned for evaluation.The evaluation was performed by the supplier and could confirm the customer complaint for the lens did not attach properly.A visual inspection was performed and showed the scope to have distal tip damage, a dented outertube and broken lenses.This damage is caused by contact with another source.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.No manufacturing related defects were observed.Correction in report source: foreign should be marked.Other specify: costa rica.
|
| |
|