| Model Number |
R SERIES |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
10/01/2019
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device was unable to obtain an ecg signal.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's report was observed and attributed to a faulty integrated circuit on the analog board.The device was recertified and returned to the customer.Please refer to the medwatch report 1220908-2019-03192 for a medwatch report inadvertently filed for this same patient event.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|