| Catalog Number |
03P84-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
11/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).The investigation was completed on 12/11/2019.A review of the device history record confirmed the lot passed finished goods release criteria.Retained cartridge testing met the acceptance criteria found in q04.01.003 rev.Ad, appendix 1- product complaint level 2 and level 3 investigation procedure.No deficiency has been identified.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, abbott point of care was contacted by a customer regarding i-stat crea cartridges that yielded a suspected discrepant creatinine results on a patient.There was additional patient information available at the time of this report.Method: i-stat, creatinine: 1.9 mg/dl.Cobas, 0.9 mg/d.Test/collection times: unk.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Search Alerts/Recalls
|