| Catalog Number |
QD11-S |
| Medical Device Problem Code |
Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Udi: (b)(4).The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported from (b)(6) that during service and evaluation, it was determined that the attachment device had a worn bearing.It was further determined that the device failed pretest for temperature assessment.It was noted in the service order that pre-operatively the device was not working properly.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|