• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC 10.5CM ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC 10.5CM ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT Back to Search Results
Catalog Number QD11-S
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Udi: (b)(4).The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Event or Problem Description
It was reported from (b)(6) that during service and evaluation, it was determined that the attachment device had a worn bearing.It was further determined that the device failed pretest for temperature assessment.It was noted in the service order that pre-operatively the device was not working properly.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10.5CM ANGLE ATTACHMENT
Common Device Name
MOTOR, DRILL, ELECTRIC - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
MDR Report Key9378410
Report Number1045834-2019-54988
Device Sequence Number10899219
Product Code HBC
UDI-Device Identifier00845384006381
UDI-Public00845384006381
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K011444
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberQD11-S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/12/2019
Initial Report FDA Received Date11/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-