|
Reported event: -customer reported that the silver end of square driver came loose from the blue handle while taking apart arrays/array clamps.Device evaluation and results: device evaluation was performed and the square screwdriver event was confirmed.Device history review: review of the device history records indicate 379 devices were manufactured and inspected on 07-02-2018 with no reported discrepancies.Complaint history review: based on the device identification, the catsweb and trackwise complaint databases were reviewed from 2011 to present for similar reported events regarding the square screwdriver.There have been 50 other events for the referenced lot number.Prs # 1914925, 1919787, 1935822, 1943690, 1947954, 1953514, 1967031, 1970724, 1987510, 1978669, 1995528, 2001610, 2017816, 2018871, 2019846, 2022521, 2028928, 2030818, 2031968, 2036455, 2045046, 2045740, 2051183, 2073709, 2082053, 2103869, 2114279, 2114688, 2117018, 2122244, 2125456, 2127852, 2139397, 2149504, 215096, 2154940, 21666790, 2188146, 2188148, 2192078, 2204673, 2208809, 2209518, 2210264 , 2210272, 2211307, 2213526, 2221304, 2221688 and 2228967 - were found to be from the same lot as the part in this complaint.The parts from these prs displayed the same failure.Conclusions: the square screwdriver showed the masterbond that holds the shaft and handle cracked.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated that capa 1694870 was initiated for this part number.
|