• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US LPS UNIV TIB HIN INS XXSM 23MM; LPS AND S-ROM : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US LPS UNIV TIB HIN INS XXSM 23MM; LPS AND S-ROM : KNEE TIBIAL INSERT Back to Search Results
Catalog Number 198727023
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); No Code Available (3191)
Date of Event 11/01/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the distal femoral components removed due to infection.Lps poly and hinge pin that were placed on (b)(6) 2019 removed, tibial and patellar components removed that were placed (b)(6) 2017.Doi: (b)(6) 2019 (femoral components); doi: (b)(6) 2019 (lps poly and hinge pin); doi: (b)(6) 2017 (tibial and patella components); dor: (b)(6) 2019.Right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LPS UNIV TIB HIN INS XXSM 23MM
Common Device Name
LPS AND S-ROM : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9379217
Report Number1818910-2019-118228
Device Sequence Number2107343
Product Code KRO
UDI-Device Identifier10603295079385
UDI-Public10603295079385
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K091453
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number198727023
Device Lot NumberJ2644D
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/01/2019
Supplement Date Received by Manufacturer12/18/2019
Initial Report FDA Received Date11/26/2019
Supplement Report FDA Received Date12/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-