The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The lab analysis concluded that the device fractured by the initiation and subsequent propagation of fatigue cracking.The fatigue cracking eventually propagated to an extent that the device could not bear the imposed patient loading.Fatigue cracking is caused by the plate bearing cyclic stresses in excess of the material endurance limit over a period of time.These stresses may be caused by excessive patient activity prior to full bone union, loads in excess of the material¿s strength, and/or poor bone quality.The medical investigation concluded that no clinical/medical documents were provided for this complaint.The root cause of the breakage cannot be definitively concluded, however excessive stress prior to full bone union cannot be ruled out.Without supporting medical documents, a thorough investigation cannot be performed.Should information become available, this complaint can be re-assessed.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.Some potential probable causes of the event could include material overloading or overstressing.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|